HAD - V4.0 Filter integrity tester Online/Offline

FOB Reference Price:Get Latest Price
US $6,500-7,200 / Sets | 1 Set/Sets (Min. Order)
Report Suspicious Activity
Overview
Quick Details
Place of Origin:
Shanxi, China (Mainland)
Brand Name:
YM-NY
Model Number:
INV4NY04
Power:
Electronic
Usage:
Auto Testing Machine
Power supply:
100-240 VAC, 50HZ, 120W
Operation pressure:
100-8000mbar (115psi)
Temperature range:
+ 5 ℃ ~ +40 ℃
Relative humidity:
10-80%
Print Function:
Built-in printer
Historical Records:
500 sets
Display Screen:
5.7 TFT,Single color
Product number:
HZYJbrwl012
SupplyAbility
SupplyAbility:
10 Set/Sets per Month
Packaging & Delivery
Packaging Details
Packaging Details: padded storage carton
Delivery Detail: about 10 days
Port
shenzhen
Lead Time :
Shipped in 10 days after payment
Product Description

Features

1. Automatic air inlet design, to ensure efficiency and test accuracy at the same time;

2. The instrument actually measures the up stream volume, so the test result is more accurate;

3. Domestic first model to introduce water immersion method;

4. Widely used by customers, with a high market share and cost performance;

5. Automatic control of air intake level, with a smooth and safe air intake, to ensure the service life of filter element;



Function

Manual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test;Diffusion Flow Test;Water Immersion Test

Test Accuracy

Upstream Volume:± 4%;Bubble Point:± 50mbar ;Diffusion Flow Test:± 4% ;Water Immersion Test:± 0.01ml

Test Scope

Bubble Point :590-6900mbar Diffusion Flow Speed:1-180ml/min WI:0.01-30ml/min

Parameters

Power Supply 100 – 240 VAC, 50HZ, 120W Spare battery(Selection)

Operation Pressure 100-10000 mbar (145psi) ;

Unit mbar

Operation Condition Ambient Temperature:+5℃ ~ +40℃;Relative Humidity:10-80%

Dimension(mm) 400(Depth)*240(Width)*270(Height1)*220(Height2)

Function Manual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test; Pressure Holding Test;Diffusion FLow Test;Water Immersion Test;Ultrafitration Membrane Test

Test Accuracy Upstream Volume Test:± 4%;Bubble Point Test:± 50mbar ;Diffusion Flow Test:± 4% ;Water Immersion:± 0.01ml

Test Scope BP:100-8000mbar DF:1-600ml/min WI:0.01-100ml/min

Applied Range Symmetric and asymmetric membrane test, needle filter, filter bag, flat filter, filter cartridge (9 core within 20 inches), ultrafiltration membrane, ultrafiltration column, a variety of irregular filters

Authority Management User Name and Password to Login ,4 Authority Grades,meeting the FDA 21CFR PART 11requirements

Reservation Solution 80 sets

Print Function Built-in printer;

Historical Records More than 5000 sets unlimited sets records

Records Copy Support USB disk to export (Include test curve)

Display Screen 7 inch, color touch screen

Serial Connection RS232/USB

Language Chinese/English

Work Mode Online/Offline

Applicable Environment Above B level

Weight 7.5KG


Detailed Images

Selection guide

Packing & Delivery